  • Material: ABS
  • Model Number: H300S
  • Origin: Mainland China
  • Item Type: Braces & Supports
  • Monitor Size: 4 inch
  • Parameters: SPO2,NIBP,TEMP,ECG
  • Application: Neonate,Adult,Child
SPO2+ECG+NIBP+TEMP+

Display:4” Color TFT

Dimension: 146mm*67mm*30mm

Weight: 250g with rechargeable battery

Temperature

  Operating 5℃~40℃

Storage/Transportation -20℃~65℃

Humidity

Operating             30% to 75% noncondensing

Storage                     ≤80% noncondensing

Altitude

Operating Altitude     Up to 5,000 meters

Hyperbaric Pressure         Up to 405.3 kPa

Main unit

Displays

  Measurement results     LED, red

  Numeric Displays                 7-segment LEDs, red

  Low Battery             LED, red

ECG

Lead  7 lead (R, L, F, N, C1 or RA,LA, LL, RL, V1)

Lead mode     I, II, III, aVR, aVL, aVF, V1, CAL

Wave     3waves

Gain selection´0.25V/mV, ´0.5V/mV, ´1V/mV,´2V/mV

Heart  Rate:

Range

Adult15 ~ 300bpm

Neo/Ped15 ~ 350 bpm

Accuracy±1%or ±1bpmBigger

Resolution   1 bpm

Sensitivity   > 200 uV(peak to peak)

Bandwidth

Diagnositic0.05 ~100 Hz

Monitor0.5 ~ 40Hz

Surgery        1 ~ 20 Hz

Stronger Filter5 ~ 20Hz

NIBP

Measuring Technology:                  Automatic oscillating measurement

Mode:                  Manual, Auto, STAT

Measuring Interval in AUTO Mode :       1 ~ 90 (Min)

Measuring Interval in Continuous Mode:   5 (Min)

Pulse Rate Range:              40 ~ 240 (bpm)

Alarm:              SYS, DIA, MEAN

Measuring Range

 Adult Mode

SYS 40 ~270 (mmHg)

MEAN      20 ~230(mmHg)

DIA 10 ~210 (mmHg)

          Pediatric Mode

            SYS     40 ~ 200 (mmHg)

            MEAN 10 ~ 150 (mmHg)

            DIA     20 ~ 165 (mmHg)

 Neonatal Mode

            SYS 40 ~ 135 (mmHg)

            MEAN 20 ~ 105 (mmHg)

            DIA 10 ~ 95 (mmHg)

     Resolution

       Pressure 1mmHg

     Accuracy

Pressure

Maximum Mean error

±5mmHg

Maximum Standard deviation

8mmHg

Overpressure Protection

           Adult Mode 290(mmHg)

       Pediatric Mode        240 (mmHg)

Neonatal Mode145 (mmHg)

Alarm Limit Setting

SYS            40~280 mmHg,

DIA            10~220 mmHg,

SPO2

Measurement Range      

                          Spo2: 0~100%

PR:0-254bpm

Perfusion Index:             0.025%-20%

  Accuracy range:

               SpO2:70%-100%

               PR:30-254bpm

                       Perfusion Index: 0.05%-20%

Measurement accuracy

SpO2 saturation:

Adult (including pediatric ) + 2digit(70-100%)       not defined (<70%)

Neonate  + 3 digits     not defined(<70%)

On motion condition + 3digit

Pulse rate:

Adult (including pediatric) + 3digits

Neonate  + 3 digits

On motion condition + 3digit

Alarm Range:0%~100%

Temperature

    Range:25 — 45°C

    Resolution:    0.1°C

    Sensitivity:±0.1°C

Power Source

External AC power or internal battery

AC power:

100-240VAC, 50/60Hz, 30VA

Fuse(self-recovery):

Input fuse: 2A/250V

Fuse (battery):60Vdc/3A(max.)

Battery

Lithium ion rechargeable batteries: 3.6V/4.2Ah

 