ECG Holter 12 Channel 12 Lead EKG System 24 Hours Multi Language Monitoring PC SW CE FDA

ECG Holter 12 Channel 12 Lead EKG System 24 Hours Multi Language Monitoring PC SW CE FDA

$511.54

Add to Wishlist
Add to Wishlist
SKU: CJZN1707211 Category:
Guaranteed Safe Checkout

Introduction

Dynamic ECG Systems adopt 12-lead system of international standard, which can continuously record ECG waveform for 24-hour and analyze ECG waveform by the PC software. It is applicable for use in medical institution and community.

Function

1)The recorder can real-time store ECG data for 24-hour.

2)Arrhythmia analysis basing on MCSSTM and TMCATM can greatly reduce doctor’s work.

3)Sync analysis for 12-lead, which ensures that QRS waveform can be found accurately, and no distortion.

4)With more than 10 templates(such as atrial premature beat, ventricular premature beat, long interval, atrial fibrillation, etc.) and many user-defined modules, which can almost distinguish any kind of pathologic waveforms.

5)Flexible analysis channel selection function, which ensures that any channels can be selected as the main analysis channel.

6)Flexible atrial fibrillation analysis, which ensures that physicians can use the whole or segmented AUTO/manual analysis, more accurate in atrial fibrillation analysis.

7)All pacemakers(such as AAI, VVI, DDD etc.) can be analyzed by the powerful pacemaker analysis function.

8)With many pacing analysis templates, such as “Dual chamber pacing”, “Atrial pacing”, “Ventricular pacing”, “Ventricular Pseudofusion” and “Ventricular asynchronous pacing”, etc.

9)Single or full-lead ECG in any time bucket can be reviewed by the fast review analysis function.

10)With analysis functions for 5-minute, 1-hour and 24-hour heart rate variability.

11)One-key print, convenient and fast to print the reports.

12)Perfect case management function.

13)Abnormal ST can be analyzed according to the whole and event by the advanced ST segment analysis and total load of myocardial ischemia, which can help the doctors comprehensively judge myocardial ischemia.

14)The risk of sleep breath pause can be predicted by the unique “Sleep breath pause syndrome”analysis function.

15)The death risk in patients with myocardial infarction can be predicted by the “HRT” analysis function.

16)“T-Wave alternation” is an important index to predict the malignant arrhythmia and sudden cardiac death.

17)With QTD(QT dispersion), VCG(vector cardiogram), VLP(ventricular late potential) and TVCG(time vector cardiogram) analysis modules, more valuable in analysis report.

Performance

Lead: standard 12-lead

Record time: 24-hour

Power supply: two “AA” batteries

Interface: USB2.0

Sampling accuracy: 24bit

Sampling frequency: 250Hz

Dynamic range: 10mV

Input impedance: ≥10MΩ

CMR: ≥60dB

Gain accuracy: At all possible gain settings, the maximum amplitude error of the output is ±10%

Gain stability: One minute after energizing the device, the gain change in stable ambient conditions shall not exceed 3% over a 24h period.

System noise: ≤50µVp-p

Multichannel crosstalk: ≤0.2mV

Frequency response: The amplitude response to sinusoidal signals within the frequency range 0.05Hz to 40Hz shall be between the +3dB to -3dB of the response at 5 Hz.

Minimum measure signal: 50 µV

With the function in the presence of pacemaker pulses

Timing accuracy: the overall error during 24h shall not exceed 30s.

Accessories

Lead cable (1 set)

ECG electrode (1 bag)

USB cable(1)

Bag(1)

User manual(1)

Physical characteristic

Dimension: 111mm(L) × 61mm(W) × 25mm(H)

Weight: about 105g(without batteries)

Optional configuration

Refer to BOM.

Certification

CE, FDA

Weight 0.70 kg
Dimensions 206 × 197 × 105 cm
Model

TLC5000

Reviews

There are no reviews yet.

Be the first to review “ECG Holter 12 Channel 12 Lead EKG System 24 Hours Multi Language Monitoring PC SW CE FDA”
Review now to get coupon!

Your email address will not be published. Required fields are marked *

ECG Holter 12 Channel 12 Lead EKG System 24 Hours Multi Language Monitoring PC SW CE FDA
Scroll to Top
0