Introduction
 
 ECG100G is a kind of single channel electrocardiograph, which features in printing ECG waveform by thermal printing system, prompting for “Lead off” or “Lack of paper”, multi-language interface, recording ECG waveform in AUTO/Manual mode, convenient to operate.
 Function
 
 1)Adopt digital signal processing technology and get high-quality ECG waveform via power frequency filter, baseline filter and EMG filter of ECG signal.
 
 2)Adopt high-resolution thermal printing system(8 dot/mm), no need for any adjustments. Recording frequency: up to 150Hz.
 
 3)Real-time and continuously record clear and accurate ECG waveform and annotation character(including lead mark, sensitivity, paper speed and filter state, etc.).
 
 4)In AUTO mode, finish recording by pressing the button once, which improves work efficiency.
 
 5)Multi-language (Chinese, English, French, Italian, German, Spanish, Turkish, Russian and Kazakh) interface and report.
 
 6)In optimal DC state, up to 7-hour standby time, continuous print up to 4-hour, record up to 150 pieces of ECG waveform.
 Performance
 
 Input mode: floating and defibrillation protection
 
 Frequency response:
 
 Rated input amplitude Input frequency and waveform Relative output response
 
 1.0 0.67Hz~40Hz, Sine wave ±10%a
 
 0.5 40Hz~100Hz, Sine wave +10 %, -30 %a
 
 0.25 100Hz~150Hz, Sine wave +10 %, -30 %a
 
 0.5 150 Hz ~ 500 Hz, Sine wave +10 %, -100 %a
 
 1.5 ≤1Hz,200ms, Triangle wave +0 %, -10 %b
 
 a relative to 10Hz b relative to 200 ms
 
 Time constant:≥ 3.2s
 
 CMRR: >105dB
 
 Filter: power frequency(AC50/60 Hz), EMG(25 Hz/35 Hz (-3 dB)), baseline drift filter
 
 Recording way: thermal printing system
 
 Specification of recording paper: 50 mm(W)×20 m(L) high-speed thermal paper
 
 Time base selection(paper speed): 6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s, error: ±5%
 
 Gain control(sensitivity): 2.5, 5, 10, 20 mm/mV, accuracy is ±2%; Standard sensitivity: 10 mm/mV±0.2 mm/mV
 
 Recording mode: auto mode and manual mode
 
 Safety classification: Class I, type CF defibrillation-proof applied part
 
 Polarization resistance voltage: ±610mV
 
 Noise level: ≤12 µVp-p
 
 ECG signal input sampling frequency: 32 kHz
 
 Waveform data processing sampling frequency: 1 kHz
 
 Sampling accuracy: 24-bit
 
 The minimum detection signal: 10 Hz, 20 µV(peak-peak value) deflected sinusoidal signal can be detected
 
 Accuracy of input signal: ±5%
 
 Amplitude quantization: ≤5µV/LSB
 
 Interchannel time deviation: <100 µs
 
 Power supply:
 
 AC: 100V~240V(50/60Hz)
 
 DC: 7.4V/2000mAh rechargeable lithium battery
 
 Degree of protection against ingress of liquid: IPX0
 
 Working mode: continuous working
 Accessories
 
 Standard configuration:
 
 A lead cable
 
 A limb electrode
 
 A chest electrode
 
 A thermal recording paper
 
 A power cord
 
 An earth wire
 
 A user manual
 Physical characteristic
 
 Working environment
 
 Temperature: 5℃~40℃
 
 Relative humidity: 25%~95%(non-condensing)
 
 Atmospheric pressure: 700 hPa~1060 hPa
 
 Transport and storage environment
 
 Temperature: -20 ℃~+55 ℃
 
 Relative humidity: ≤95%
 
 Atmospheric pressure: 500 hPa~1060 hPa
 
 Dimension: 315mm(L)×215mm(W)×77mm(H)
 
 Net weight: 1.5Kg
 Certification
 
 SFDA, CMC, CE and FDA.
 
 
															







 



 
															





Reviews
There are no reviews yet.