- Brand Name: LanYuXuan
- Origin: Mainland China
- Certification: CE
- Type: Blood Analysis SYstem
- Brand Name: FORERMED
- Model Number: BC-5000
- Place of Origin: Henan, China
- Instrument classification: Class II
- Warranty: 1 Year
- After-sale Service: Online technical support
- Product name: BC-5000 Auto Hematology Analyzer
- Part: 5-part hematology analyzers
- Display: 10.4 inch large TFT touch screen with user-friendly software
- Large storage capacity: up to 20,000 samples
- Throughput: 40 samples per hour
- Sample volume: 15μL
- Application: clinical labs
- Multi-language: Chinese, English
- Parameters: 23 parameters
- Communication: LAN port supports HL7 protocol
Whole blood mode, Capillary whole blood mode and Prediluted mode
Tri-angle Laser scatter + Chemical dye + Flow cytometry technology
Dedicated optical counting channel for Basophil measurement
Powerful capability of flagging abnormal cells
Large storage capacity: up to 20,000 samples
Throughput: 40 samples per hour
Sample volume is only 15μL which is ideal for pediatrics
Compact
Two kinds of lyse reagents are located inside of BC-5000, which helps
the small labs to save space.
Utility
convenience to clinicians. Users can complete all instrument operations
on the screen, practically eliminating the need for an external PC.
Technology
-conductor laser has smaller size, lower cost and longer life cycle.
Effiffifficient
and also less consumed by BC-5000. Original QC and calibrator are also
provided to ensure the hematology analyzer’s traceability and testing
quality.
convenient for the users in children’s hospitals, etc. For Prediluted mode,
BC-5000 has higher dilution ratio than other 5-part hematology
analyzers, thus it brings a better mixing effffect.
BC-5000 users to transmit data conveniently and connect with printers,
keyboard, mouse, barcode scanner, etc.
Improved DC impedance technology is used to count and size the RBC
and PLT. The smaller counting aperture (50 μm in diameter) provides
better performance on samples with low PLT.
Technical Specififications
Impedance method for RBC and PLT counting
Cyanide free reagent for hemoglobin test
Flow Cytometry (FCM) + Tri-angle laser scatter + Chemical dye
method for WBC 5-part difffferential analysis and WBC counting
Parameters
23 parameters: WBC, Lym%, Mon%, Neu%, Bas%, Eos%, Lym#,
Mon#, Neu#, Eos#, Bas#, RBC, HGB, HCT, MCV, MCH, MCHC,
RDW-CV, RDW-SD, PLT, MPV, PDW, PCT
3 histograms for WBC, RBC and PLT
3 scattergrams for WBC difffferential
Reagent
Diluent, DIFF lyse, LH lyse, probe cleanser
Performance
Parameter Linearity Range Precision Carryover
WBC 0-100×109 /L ≤2% (4-15×109 /L) ≤0.5%
RBC 0-8×10¹²/L ≤1.5% (3.5-6.0×10¹²/L) ≤0.5%
HGB 0-250g/L ≤1.5% (110-180g/L) ≤0.6%
PLT 0-1000×109 /L ≤4.0% (100-500×109 /L) ≤1.0%
Throughput
40 samples per hour
Prediluted mode 20 μL
Whole blood mode 15 μL
Capillary whole blood mode 15 μL
10.4 inch TFT Touch Screen
Multi-language
Chinese, English
Data Storage Capacity
Up to 20,000 results including numeric and graphical
information
Communication
LAN port supports HL7 protocol
Interface
USB, LAN
Support uni-directional LIS
Printout
External Laser printer / Inkjet printer, various printout formats
and user-defifined formats
Operating Environment
Temperature: 10°C~30°C
Humidity: 20%~85%
Air pressure:70 kPa~106 kPa
Power requirement
|
100V-240V 50Hz/60Hz
|
||||||
Dimension and Weight
|
Depth(400 mm) x width(320 mm) x height(410 mm)
|
||||||
Weight
|
24kg
|
Reviews
There are no reviews yet.